Neurology · 510(k) · 2023
Alignment System Cranial, Alignment Software Cranial, Cirq Alignment Software Cranial Biopsy, Cirq Alignment Software Cranial sEEG, Varioguide Alignment Software Cranial
Brainlab AG. Cleared by FDA on 9 September 2023 under submission K223864, product code HAW (neurological stereotaxic instrument). What FDA's record says, in FDA's words.
Device record · K223864
- Device. Alignment System Cranial, Alignment Software Cranial, Cirq Alignment Software Cranial Biopsy, Cirq Alignment Software Cranial sEEG, Varioguide Alignment Software CranialDEVICE
- Company. Brainlab AG (DE)COMPANY
- Decision. 9 September 2023, 510(k), substantially equivalent (traditional review); received 23 December 2022DECISION
- Panel. NeurologyPANEL
- Product code. HAW, Neurological Stereotaxic InstrumentCODE
- Regulation. 21 CFR 882.4560, Class IIREG
- Source. FDA database entry · decision document (10 pages)SOURCE
Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.
What FDA cleared it for
The Indications for Use statement, verbatim, from the decision document.
The Stealth Autoguide System is a positioning and guidance system intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments, based on a pre - operative plan and feedback from an image -guided navigation system with three- dimensional imaging software. The Stealth Autoguide System is a remotely -operated positioning and guidance system, indicated for any neurological condition in which the use of stereotactic surgery may be appropriate (for example, stereotactic biopsy, stereotactic EEG, laser tissue ablation, etc.). The Cirq Robotic Alignment Module is an intraoperative robotic positioning system and an accessory to compatible Brainlab IGS Cranial software applications. Using spatial information from an image guided navigation system, the Cirq Robotic Alignment Module enables the surgeon to align and hold surgical instruments according to pre -planned trajectories. The Cirq Robotic Alignment Module is indicated for stereotactic biopsies of intracranial lesions The Cranial IGS System, when used with a compatible navigation plat form and compatible instrument accessories, is intended as an image-guided planning and navigation system to enable navigated surgery. It links instruments to a virtual computer image space on patient image data that is being processed by the navigation platform. The system is indicated for any medical condition in which a reference to a rigid […]
What the statement says
Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.
- Mentionsnone of the tracked terms matched
Adverse-event reports in MAUDE, by product code
As of , MAUDE held 42,962 reports filed against devices carrying product code HAW. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:
30 September 2026 · Malfunction · StealthStation¿ S8 system · MEDTRONIC NAVIGATION, INC. · report 1723170-2026-01386
Medtronic received information regarding a Navigation system being used outside of a procedure. It was reported that an inaccuracy was experienced while working with 2 different instruments in the same case. The application team found that the issue was with the end user (human issue). After that they verified the accuracy using demo sample. It was reported that all instruments were accurate and that the issue was with the end user. The amount of the inaccuracy was 10 mm, but it was not […]
30 September 2026 · Malfunction · StealthStation¿ S8 system · MEDTRONIC NAVIGATION, INC. · report 1723170-2026-01389
Medtronic received information regarding a navigation system being used outside of a procedure. It was reported that the system had wear on the main power cable to the system. The grey coating was changing to white. There were no tears and no cable showing. There was no patient involvement.
30 September 2026 · Malfunction · StealthStation¿ S8 system · MEDTRONIC NAVIGATION, INC. · report 1723170-2026-01390
30 September 2026 · Malfunction · brand not stated · MEDTRONIC NAVIGATION, INC. · report 1723170-2026-01391
29 September 2026 · Malfunction · StealthStation¿ S8 system · MEDTRONIC NAVIGATION, INC. · report 1723170-2026-01382