Radiology · 510(k) · 2020

Auto Positioning

GE Hangwei Medical Systems, Co., Ltd.. Cleared by FDA on 16 January 2020 under submission K192956, product code JAK (system, x-ray, tomography, computed). What FDA's record says, in FDA's words.

Device record · K192956

  1. Device. Auto PositioningDEVICE
  2. Company. GE Hangwei Medical Systems, Co., Ltd. (CN)COMPANY
  3. Decision. 16 January 2020, 510(k), substantially equivalent (traditional review); received 21 October 2019DECISION
  4. Panel. RadiologyPANEL
  5. Product code. JAK, System, X-Ray, Tomography, ComputedCODE
  6. Regulation. 21 CFR 892.1750, Class IIREG
  7. Source. FDA database entry · decision document (12 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

The Auto Positioning feature provides an alternate, streamlined and efficient workflow and performs quality and safety checks for the CT technologist in setting up CT examinations up to the initiation of the first scout scan. Auto Positioning acquires 3D spatial information of the individual patient on the table and combines it with information from the selected protocol to automatically calculate and visually display the scout’s start and end locations. Concurrently it checks for proper patient orientation, determines the table height for optimum patient centering, and checks for potential contact between the patient and the gantry. Upon acceptance by the technologist the patient is automatically moved to the correct scout start location. Use of Auto Positioning is intended to provide consistent patient positioning for optimal image quality and automatic exposure control.

K192956, FDA decision document · source

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 5,281 reports filed against devices carrying product code JAK. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

25 September 2026 · Malfunction · REVOLUTION EVO · GE MEDICAL SYSTEMS, LLC · report 50896044

The GE CT table wouldn't move during stroke. Patient had to be physically moved to resume scan, delay in care. GE CT table has had the same issues with the table working intermittently x 3 weeks. Multiple techs have worked on it and the same problem persists.

25 September 2026 · Malfunction · Needle Guide · Quantum Surgical SAS · report 3023140882-2026-00002

23 September 2026 · Malfunction · ILLUMISITE · COVIDIEN LP - SUPERDIMENSION INC · report 3004962788-2026-00087

It was reported that during the procedure, one locatable guide indicated as outside of the sensor sensing volume and another locatable guide had difficulty retracting from the extended working channel. The physician completed the superDimension portion of the case. No patient complications have been reported as a result of this event.

23 September 2026 · Other · NL4000 · Samsung HME America, Inc. · report 3004938766-2026-00043

23 September 2026 · Malfunction · ILLUMISITE · COVIDIEN LP - SUPERDIMENSION INC · report 3004962788-2026-00088

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.

More devices under product code JAK, in Radiology, from GE Hangwei Medical Systems, Co., Ltd., or cleared in 2020. The whole record is at FDA AI-enabled medical devices.

A device's authorized use is a sentence the record can be read against. What Apodicta does on a medical record is that reading: every line of a note, a claim or a report proved to its source or held, before it renders.

Show me on one of my files →

This device is authorized for a stated use. See a record read against what it must prove, on your own file.

Show me on my files