Radiology · 510(k) · 2023

HepaFatSmart (V2.0.0)

Resonance Health Analysis Services Pty Ltd. Cleared by FDA on 20 June 2023 under submission K231459, product code LNH (system, nuclear magnetic resonance imaging). What FDA's record says, in FDA's words.

Device record · K231459

  1. Device. HepaFatSmart (V2.0.0)DEVICE
  2. Company. Resonance Health Analysis Services Pty Ltd (AU)COMPANY
  3. Decision. 20 June 2023, 510(k), substantially equivalent (special review); received 19 May 2023DECISION
  4. Panel. RadiologyPANEL
  5. Product code. LNH, System, Nuclear Magnetic Resonance ImagingCODE
  6. Regulation. 21 CFR 892.1000, Class IIREG
  7. Source. FDA database entry · decision document (14 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

Intended use: HepaFatSmart is intended for the quantitative measurement of volumetric liver fat fraction (VLFF), proton density fat fraction (PDFF) and steatosis grading. HepaFatSmart is an application that is used for the non-invasive evaluation of liver tissue by utilising magnetic resonance images to evaluate the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton density signal in the liver. Indications for use: Support clinical diagnoses in individuals with confirmed or suspected fatty liver disease; Support the subsequent clinical decision making processes for patients under management for fatty liver related disease or metabolic syndromes; Aid in the assessment and screening of living donors for liver transplant. Results, when interpreted by a trained physician can be used to support clinical diagnoses about the status of liver fat content, the subsequent clinical decision making processes for the management of fatty liver related diseases, metabolic syndromes, liver donor screening and lifestyle change. HepaFatSmart can be used to analyse the MRI images of patients of all populations independent of age and gender, with suspected clinical conditions related to the level of liver fat.

K231459, FDA decision document · source

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 4,787 reports filed against devices carrying product code LNH. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

29 September 2026 · Malfunction · Achieva 1.5T Dstream · Philips Medical Systems Nederland B.V. · report 3042177665-2026-000174

28 September 2026 · Malfunction · Achieva 1.5T · Philips Medical Systems Nederland B.V. · report 3042177665-2026-000166

Philips received a report that the Quadrature Body Coil (QBC) seal on the front cover is coming loose. This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the Magnet covers.If this seal is loose, it is likely that this could lead to serious injury.

28 September 2026 · Malfunction · Ingenia · Philips Medical Systems Nederland B.V. · report 3042177665-2026-000167

Philips received a report that the upper Quadrature Body Coil (QBC) seal has peeling off. This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the Magnet covers.If this seal is peeled it is likely that this could lead to serious injury.

28 September 2026 · Malfunction · Ingenia · Philips Medical Systems Nederland B.V. · report 3042177665-2026-000168

Philips received a report that the Quadrature Body Coil (QBC) seal has come loose. This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the Magnet covers.If this seal comes loose it is likely that this could lead to serious injury.

28 September 2026 · Malfunction · Ingenia · Philips Medical Systems Nederland B.V. · report 3042177665-2026-000169

Philips received a report that the upper Quadrature Body Coil (QBC) seal has come loose This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the Magnet covers.If this seal is loose it is likely that this could lead to serious injury.

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.

More devices under product code LNH, in Radiology, from Resonance Health Analysis Services Pty Ltd, or cleared in 2023. The whole record is at FDA AI-enabled medical devices.

A device's authorized use is a sentence the record can be read against. What Apodicta does on a medical record is that reading: every line of a note, a claim or a report proved to its source or held, before it renders.

Show me on one of my files →

This device is authorized for a stated use. See a record read against what it must prove, on your own file.

Show me on my files