Radiology · 510(k) · 2021
qXR-BT
Qure.ai Technologies. Cleared by FDA on 21 December 2021 under submission K212690, product code QIH (automated radiological image processing software). What FDA's record says, in FDA's words.
Device record · K212690
- Device. qXR-BTDEVICE
- Company. Qure.ai Technologies (IN)COMPANY
- Decision. 21 December 2021, 510(k), substantially equivalent (traditional review); received 25 August 2021DECISION
- Panel. RadiologyPANEL
- Product code. QIH, Automated Radiological Image Processing SoftwareCODE
- Regulation. 21 CFR 892.2050, Class IIREG
- Source. FDA database entry · decision document (8 pages)SOURCE
Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.
What FDA cleared it for
The Indications for Use statement, verbatim, from the decision document.
The qXR-BT device is intended to generate a secondary digital chest X-ray image that facilitates confirmation of the position of a breathing tube and an anatomical landmark on adult chest X- rays. This device is intended for use by licensed physicians who are trained in the evaluation of breathing tube placement on chest X-rays. The qXR-BT image provides adjunctive information and is not a substitute for the original PA/AP image. K212690 2 5 DEVICE DESCRIPTION qXR-BT is a standalone image analysis software used during the review of digital chest radiographic images, intended to facilitate determining the position of tip of the breathing tube relative to the carina. Standard of care medical imaging workflows are well established, and include pre-existing software components such as a PACS, DICOM viewer and imaging worklist; qXR-BT is designed to integrate with these components. X-rays are sent to qXR-BT by means of transmission functions within the user’s PACS system. Upon completion of processing, the qXR-BT device returns results to the user’s PACS or other user-specified radiology software system or database. The input to the qXR-BT device is a chest X-ray (AP and PA, referred to as frontal) in digital imaging and communications in medicine (DICOM) format. The qXR-BT device produces PDF and DICOM format outputs that enable users to view the position of a breathing tube and an anatomical landmark (carina). The PDF format output contains preview images that show segmented […]
The product code's definition, from FDA's classification record. To provide automated radiological image processing and analysis tools. Software implementing artificial intelligence including nonadaptive machine learning algorithms trained with clinical and/or artificial data. In these devices, the algorithm training data typically impacts device performance. Adaptive AI algorithms are not within scope of this product code.
What the statement says
Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.
- Modalities namedradiographs (X-ray)
- Mentionsan adjunct to the clinician, not a replacement
- Mentionstriage or notification
- Mentionsmeasurement or quantification
- Mentionsadult patients
Adverse-event reports in MAUDE, by product code
As of , MAUDE held 10 reports filed against devices carrying product code QIH. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:
2 July 2026 · Malfunction · SYNAPSE PACS · FUJIFILM HEALTHCARE AMERICAS CORPORATION · report 3004972322-2026-00002
ON 06/04/2026, FUJIFILM HEALTHCARE AMERICAS CORPORATION BECAME AWARE OF A BUG IN SYNAPSE PACS V7.4.310 THAT COULD POTENTIALLY LEAD TO A MISDIAGNOSIS IN THE CARDIOLOGY BASED STUDIES. IT WAS FOUND DURING REGRESSION TESTING OF THE PRODUCT THAT THE ANNOTATION TAKEN AND SAVED IN AN EARLIER VERSION OF THE PRODUCT CAN SHOW THE INCORRECT VALUE IN BOTH ZERO VIEWER (ZV) AND VX WHEN REOPENING THE STUDY IN THE NEWER VERSION. THIS ISSUE WAS INTRODUCED INTO THE PRODUCT DURING AN ENHANCEMENT TO SUPPORT […]
9 April 2026 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB SE · report 8043933-2026-00021
A PATIENT RECORD REQUIRED FOR STEREOTACTIC PLANNING COULD NOT BE FOUND DURING A FRAME-BASED STEREOTACTIC BIOPSY FOR A SUSPECTED BRAIN TUMOR. AFTER BRIEF TROUBLESHOOTING, IT BECAME CLEAR THAT THE PATIENT RECORD WAS UNINTENTIONALLY MERGED WITH ANOTHER PATIENT RECORD. THE SURGEON RESOLVED THE MERGE WITH SUPPORT FROM BRAINLAB. FURTHER INVESTIGATION BY BRAINLAB UNCOVERED FOUR MORE CASES OF INCORRECT PATIENT RECORD MERGES AT THE SAME SITE THAT WERE THEREAFTER RESOLVED. ACCORDING TO THE SURGEON […]
26 February 2026 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB SE · report 8043933-2026-00011
WHILE SEARCHING FOR A MISSING PATIENT RECORD IN THE BRAINLAB PATIENT SELECTION 6.1 APPLICATION, A CUSTOMER DISCOVERED THAT THE MISSING PATIENT RECORD HAD BEEN MERGED WITH AN UNRELATED PATIENT RECORD. THE CUSTOMER RESTORED THE MERGED PATIENT RECORD TO ITS ORIGINAL CONSTITUENT PATIENT RECORDS ON THE SAME DAY. AFTER THIS, THE MISSING PATIENT RECORD BECAME AVAILABLE IN THE PATIENT LIST. ACCORDING TO THE HOSPITAL, NO PATIENT WAS AFFECTED DUE TO THE DESCRIBED ISSUE AS NEITHER OF THE PATIENTS HAS HAD […]
13 August 2025 · Malfunction · ORIGIN DATA MANAGEMENT (VERSION 3.2, SOFTWARE COMPONENT) · BRAINLAB AG · report 8043933-2025-00058
A CUSTOMER REPORTED THAT A PATIENT RECORD DID NOT APPEAR IN THE BRAINLAB PATIENT SELECTION APPLICATION WHEN THE PATIENT'S NAME WAS SEARCHED. THE CUSTOMER FOUND THE MISSING PATIENT DATA MERGED INTO A DIFFERENT PATIENT RECORD AND WAS ABLE TO RESTORE THE PATIENT RECORDS TO THEIR ORIGINAL STATES. NO PATIENT WAS AFFECTED DUE TO THIS ISSUE.
25 June 2025 · Malfunction · SONIO DETECT · SONIO · report MW5171987
SONIO DETECT SOFTWARE AI ALGORITHM IS FAULTY AND WRONGLY LABELS FETAL STRUCTURES AND ASSOCIATES THEM WITH THE WRONG BODY PARTS. THE SYSTEM WILL FREQUENTLY DISPLAY AS VISUALIZED EVEN THOUGH IT HASN'T LEAVING PROVIDERS TO GIVE FALSE NORMALCY AND OPENING THEM UP TO LIABILITIES.
Where this fits
This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.
More devices under product code QIH, in Radiology, from Qure.ai Technologies, or cleared in 2021. The whole record is at FDA AI-enabled medical devices.
A device's authorized use is a sentence the record can be read against. What Apodicta does on a medical record is that reading: every line of a note, a claim or a report proved to its source or held, before it renders.