Radiology · 510(k) · 2021

syngo.via RT Image Suite

Siemens Medical Solutions USA Inc.. Cleared by FDA on 30 July 2021 under submission K211379, product code MUJ (system, planning, radiation therapy treatment). What FDA's record says, in FDA's words.

Device record · K211379

  1. Device. syngo.via RT Image SuiteDEVICE
  2. Company. Siemens Medical Solutions USA Inc.COMPANY
  3. Decision. 30 July 2021, 510(k), substantially equivalent (traditional review); received 4 May 2021DECISION
  4. Panel. RadiologyPANEL
  5. Product code. MUJ, System, Planning, Radiation Therapy TreatmentCODE
  6. Regulation. 21 CFR 892.5050, Class IIREG
  7. Source. FDA database entry · decision document (9 pages)SOURCE

Retrieved · the fields above are FDA's; the statement below was copied from the decision document on that day.

What FDA cleared it for

The Indications for Use statement, verbatim, from the decision document.

syngo.via RT Image Suite is a 3D and 4D image visualization, multi -modality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to view existing contours, create, edit, modify, copy contours of regions of the body, such as but not limited to, skin outline, targets and organs- at-risk. It also provides functionalities to create simple geometric treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System. The software combines the following digital image processing and visualization tools: • Multi-modality viewing and contouring of anatomical, functional, and multi-parametric images such as but not limited to CT, PET, PET/CT, MRI, Linac CBCT images • Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT) • Freehand and semi -automatic contouring of regions -of-interest on any orientation including oblique • Automated Contouring on CT images • Creation of contours on images supported by the application without prior assignment of a planning CT • Manual and semi-automatic registration using rigid and deformable registration • Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points • Supports […]

K211379, FDA decision document · source

What the statement says

Pattern matches on the Indications for Use text. Each line is a fact that the text contains the words, not a finding about the device.

Adverse-event reports in MAUDE, by product code

As of , MAUDE held 807 reports filed against devices carrying product code MUJ. That code is shared by every device of this type, not this device alone, and MAUDE reports are submitted by manufacturers, facilities and the public without FDA verification. The latest reports, in the reporter’s words:

24 September 2026 · Malfunction · MONACO · Elekta Solutions AB · report 3015232217-2026-00031

The customer reported that the SRS Cone plan export for the counter clockwise beam is not being transferred correctly to MOSAIQ.

22 September 2026 · Malfunction · RayStation · RaySearch Laboratories AB (Publ) · report 3007774465-2026-00003

A user was preparing radiotherapy treatment of the right breast for a patient who had previously received RT to the left breast. The user performed a deformable registration of the two CT scans (prior treatment and current) and mapped the dose onto the current planning CT to create a cumulative dose distribution. The user noticed that the dose in the left breast (from the prior treatment) appeared cut off as it was limited by the size of the dose grid used for the prior treatment. The patient […]

22 September 2026 · Malfunction · RayStation · RaySearch Laboratories AB (Publ) · report 3010034862-2026-00003

A user was preparing radiotherapy treatment of the right breast for a patient who had previously received RT to the left breast. The user performed a deformable registration of the two CT scans (prior treatment and current) and mapped the dose onto the current planning CT to create a cumulative dose distribution. The user noticed that the dose in the left breast (from the prior treatment) appeared cut off as it was limited by the size of the dose grid used for the prior treatment. The patient […]

15 July 2026 · Malfunction · PLANONE 1 · COSYLAB, JSC, CONTROL SYSTEM LABORATORY · report 3014425747-2026-00001

DURING TREATMENT PLAN PREPARATION FOR A PATIENT, IT WAS DISCOVERED THAT COUCH SHIFTS (TREATMENT COUCH TRANSLATIONS REQUIRED TO MOVE THE PATIENT FROM THE REFERENCE/SCANNED POSITION TO THE TREATMENT POSITION) DISPLAYED IN PLANONE DIFFERED FROM THOSE DISPLAYED IN ARIA/ECLIPSE WHEN A PLAN CREATED IN PLANONE WAS TRANSFERRED TO ARIA FOR SCHEDULING. NO PATIENT HARM OCCURRED, AS THE INCONSISTENCY WAS IDENTIFIED DURING THE PLAN PREPARATION PHASE BEFORE TREATMENT DELIVERY. INVESTIGATION IDENTIFIED THE […]

2 June 2026 · Malfunction · MONACO · ELEKTA SOLUTIONS AB · report 3015232217-2026-00016

THE CUSTOMER REPORTED THAT THE NUMBERING OF THE FIRST MRI IN MONACO ALWAYS STARTS WITH MR2, AND ONLY THE SECOND MRI THEN BEARS THE NUMBER MR1.

Where this fits

This page is a record, not medical advice and not a safety rating. It copies FDA's fields and FDA's statement, and lists what the statement's text mentions. Read the labeling before relying on any of it.

More devices under product code MUJ, in Radiology, from Siemens Medical Solutions USA Inc., or cleared in 2021. The whole record is at FDA AI-enabled medical devices.

A device's authorized use is a sentence the record can be read against. What Apodicta does on a medical record is that reading: every line of a note, a claim or a report proved to its source or held, before it renders.

Show me on one of my files →

This device is authorized for a stated use. See a record read against what it must prove, on your own file.